Key Features:
- Product protection
- Operator protection
- Environmental protection
- 100% exhaust, no recirculation
- Requires connection to external duct
- Meets EN 12469
Available in sizes 1200, 1500, and 1800.
Key Features:
Available in sizes 1200, 1500, and 1800.
Technical specifications
Category | Property | Unit | 1200 | 1500 | 1800 |
Size | External dimensions excl support, WxDxH | mm | 1303×799×1405 | 1603×799×1405 | 1907×799×1405 |
External dimensions incl support, table top at 800 mm, WxDxH | mm | 1303×799×2135 | 1603×799×2135 | 1907×799×2135 | |
Internal dimensions, WxDxH | mm | 1200×520(625)×700 | 1500×520(625)×700 | 1800×520(625)×700 | |
Front window opening | mm | 200 (adjustable 160–250) | |||
Work height, basic support stand | mm | 800 or 900 (±25) | |||
Work height, adjustable support stand | mm | 700–1050 | |||
Work height, electrically elevated support stand | mm | 750–1050 | |||
Minimum door way | mm | 800 | |||
Weight | Equipment | kg | 235 | 265 | 295 |
Air Velocity | Vertical flow | m/s | 0.28 (adjustable 0.25–0.53) | ||
Inflow | m/s | 0.45 (adjustable 0.45–0.55) | |||
Deviation | ±% | <8 | |||
Noise | Noise level (EN12469) | dB(A) | 46 | 47 | 47 |
Ventilation | Down flow rate | m³/h | 680 | 840 | 1010 |
Exhaust flow rate (passive - needs ducting) | m³/h | 1070 | 1320 | 1590 | |
Filter Technology | Main/Exhaust | type | HEPA H14 EN1822, 99.999% at 0.3 µm | ||
Prefilter | type | G4 | |||
Light | LED | lux | 0–2000, dimmable | ||
Electrical Data | Voltage frequency | V/Hz | 220–240/50–60 or 110–120/50–60 | ||
Power consumption, from | W | 100 | 130 | 140 |
Dimension
Cabinet | A - Length | B - Work height | C - Top cabinet | D - Highest point | E - Work opening | W - Weight |
n-SAFE B2 1200 | 1303 mm | 700 - 1050 mm | 1953 - 2303 mm | 2042 - 2392 mm | 1200 mm | 235 kilo |
n-SAFE B2 1500 | 1603 mm | 700 - 1050 mm | 1953 - 2303 mm | 2042 - 2392 mm | 1500 mm | 265 kilo |
n-SAFE B2 1800 | 1907 mm | 700 - 1050 mm | 1953 - 2303 mm | 2042 - 2392 mm | 1800 mm | 295 kilo |
SKU | Product | Type | Language | File |
---|---|---|---|---|
Brochure | English | Download |
Dimensions | 1200, 1500, 1800 |
---|
Specialized for handling cytotoxic drugs. Features triple HEPA H14 filtration, low noise, and enhanced ergonomics. Complies with EN 12469 and DIN 12980, ideal for hazardous compounds.
This product has multiple variants. The options may be chosen on the product pageThe n-Safe Sterile is an Open Laminar Flow Cabinet, which provides a controlled environment whit vertical laminar flow. The products, which are handled inside the n-Safe Sterile are protected from any kind of contamination which makes it ideal for sterile environments.
This product has multiple variants. The options may be chosen on the product pageA highly reliable Class III cabinet, designed and manufactured according to EN 12469. Ermetic glove box for work with highly dangerous agents. Full containment, with dual HEPA filters, leak-tight valves and optional shielding.
Economical variant of the n-SAFE Class II Pro, it maintains essential security but with reduced options. Good for labs on a budget.
This product has multiple variants. The options may be chosen on the product pageThe n-Safe Class I for operator protection against low/moderate risk agents. No recirculation. Simple, robust and silent, according to EN 12469.
This product has multiple variants. The options may be chosen on the product pageLarger version of the nSafe Sterile model. Suitable for sterile work with bulky equipment and manufactured according to EN 12469. Custom Order only.
The n-Safe Pro cabinets are designed, to provide superior operator and product protection according to EN-12469 and NSF/ANSI 49 The most versatile and popular model in the range, offering high performance, ergonomic design, quiet operation, and touchscreen. Ideal for routine and research labs.
This product has multiple variants. The options may be chosen on the product pageThe n-Safe XL cabinets offer extraordinary inner dimensions, whenever the working space of a standard cabinet isn ́t sufficient: XL cabinets come with an interior height of 1000mm and an interior depth of 900 mm. The complete front can be opened to easily fit in large devices, e.g. a pipetting robot.
This product has multiple variants. The options may be chosen on the product pageReagents
N-Aminourea hydrochloride, Hydrazine carboxamide hydrochloride
Reagents
Sodium (meta)vanadate; Sodium trioxovanadate; Sodium vanadate(V); Sodium vanadium oxide; Sodium vanadium trioxide
Clean Bench and Laminar Flow
·Featuring many technologies and authoritative testing certification for reliability ·Microcomputer intelligent control panel with durable touch buttons ·Multiple safety protection functions including UV delay start ·Interlocking function to put an end to incorrect operation ·Ergonomic design to ensure comfortable operation ·Memory function to avoid repeated start-up setting ·Integrated stand with universal casters
This product has multiple variants. The options may be chosen on the product page
Sterile TPP TubeSpin® Bioreactor vessel, 50 ml, with pH and O2 sensor spots, 1 pce.
For the separation and purification of samples of cells, macromolecular proteins, blood, body fluids, DNA/RNA, plasmids, bacteria and viruses.
The ES-20 is a compact bench–top Shaker–Incubator used for mixing of biological liquids as well as for incubation and cultivation of biological liquids according to the operator set program
Applicable for storing wholeblood, medicines, biological products and other laboratory products that need to be stored at 4℃. Suitable for the storage of blood and blood articles in areas that have power shortages.
LN2 backup cooling system is an independent refrigeration system for a ULT freezer. When there is a loss of power or the temperature of the freezer rises to the high alarm set point, the LN2 backup system can be automatically activated to inject LN2 into the chamber to maintain the freezer temperature. The backup system operates on a rechargeable battery when there is a loss of main power.
LN2 backup cooling system is an independent refrigeration system for a ULT freezer. When there is a loss of power or the temperature of the freezer rises to the high alarm set point, the LN2 backup system can be automatically activated to inject LN2 into the chamber to maintain the freezer temperature. The backup system operates on a rechargeable battery when there is a loss of main power.
The ES-20 is a compact bench–top Shaker–Incubator used for mixing of biological liquids as well as for incubation and cultivation of biological liquids according to the operator set program
For the separation and purification of samples of cells, macromolecular proteins, blood, body fluids, DNA/RNA, plasmids, bacteria and viruses.
Sterile TPP TubeSpin® Bioreactor vessel, 50 ml, with pH and O2 sensor spots, 1 pce.
Applicable for blood stations, hospitals, CDC's, scientific research institutes, chemical industry and other related industries. The -30°C Biomedical Freezer can cryopreserve plasma, biological products, components, materials, and other items that need to be kept at low temperature.
This product has multiple variants. The options may be chosen on the product page
The n-Safe B2 cabinets are microbiological safety cabinets according to EN 12469. It protects the product as well as the operator and has 100% exhaust which means no recirculating air, making it ideal for volatile chemicals.
If you're representing a registered business, you may apply for a Business Account. We look forward to providing your business with a customizable Neoreax experience!